What is the best way to use indexes for multiplexing my Archer samples on Illumina platforms?

The Illumina MiSeq will work best when index diversity within a run is high. For example, if eight samples are included in a run, and the user chooses to use only one MBC Adapter paired with eight different MiSeq Index 1 Primers, the run may fail due to low barcode diversity. In this example it is best to use eight different Archer MBC Adapters paired with eight different MiSeq Index 1 Primers. Please refer to the panel specific protocols for guidance.


For Research Use Only. Not for use in diagnostic procedures. For Research Use Only. Not for use in diagnostic procedures.