What are the criteria for a sample to pass QC?

For FusionPlex RNA Assays: A sample will pass the QC filter if the average number of unique RNA start sites per control GSP2 (Gene Specific Primer 2) is 10. This metric is reported under the Read Statistics tab, labeled Average Unique Start Sites per GSP2 Control.

For VariantPlex DNA Assays: A sample will pass the QC filter if the average number of unique DNA start sites per target GSP2 is 50 or above. This metric is reported under the Read Statistics tab, labeled Average Unique Start Sites per GSP2.

This has proven to be a reasonable criterion for the success of the assay. However, users are encouraged to evaluate the results themselves and set their own cutoff.