Archer FusionPlex and VariantPlex FAQs

Why is it better to use RNA instead of DNA for fusion detection?

Why do Archer FusionPlex assays use RNA instead of DNA as input material? It all comes down to biological relevance, cost and turn-around-time. Translocations can occur anywhere in the genome, including introns and other non-coding sequences. They can also occur within the coding regions of genes with limited expression patterns. What this means is that many of the translocations that occur in a cell may not be expressed and thus have little or no biological relevance. For this reason, DNA is not the ideal substrate to search for oncogenic fusions. RNA, on the other hand, is the intermediate product of gene expression and is ideal for detecting fusions, because you are only looking at those that are expressed and potentially oncogenic. Searching for translocations in non-coding regions of the genome is time consuming and expensive. For example, DNA-based hybrid capture techniques tile over intronic regions, which can be repetitive, homopolymer-prone and span 100kb or more. This approach requires more probes, more space on your sequencer and more input material. And even then, coverage can be spotty. On the other hand, FusionPlex assays use RNA transcripts and place gene-specific primers near known fusion breakpoints, so you can identify translocations with a single primer. And because FusionPlex assays combine primers for multiple fusion targets, you can efficiently detect more fusions with less reads and input material. FusionPlex assays use RNA to detect fusions and are better, faster and cheaper than DNA-based hybrid capture techniques. Detect fusions the Archer way, with one of the many FusionPlex assays.

 


 

What values should I expect for cluster density on Illumina® sequencing runs?

The expected cluster density will vary by instrument type and kit version. Please see table below for details.

Instrument Kit Expected Cluster Density
HiSeq® High Output, TruSeq v3 750-850 k/mm2
High Output, TruSeq v4 950-1050 k/mm2
Rapid, v2 850-1000 k/mm2
MiSeq® v2 1000-1200 k/mm2
v3 1200-1400 k/mm2
MiniSeq® Mid & High Output 170-220 k/mm2
NextSeq® Mid & High Output, v2 170-220 k/mm2

Please refer to your kit-specific protocol for complete library denaturation and loading instructions.

 


 

How many Archer® samples can be multiplexed together?

The level of multiplexing depends on the number of targets and the number of reads generated by the instrument per run. This will vary for each catalog panel as well as custom panels. Custom fusion detection assays will need to be optimized to balance the number of reads needed against the level of multiplexing.

Archer® Illumina® or Ion Torrent FusionPlex Panels

Archer Illumina or Ion Torrent Panel Input Material Applications # of Genes # of Targets/Assay Recommended # of Reads
Archer FusionPlex ALK, RET, ROS1 Panel v2 RNA/TNA Fusions/SNVs 3 29 1,000,000
Archer FusionPlex Heme v2 Panel RNA/TNA Fusions 87 607 1,500,000
Archer FusionPlex NTRK Panel RNA/TNA Fusions 3 25 1,000,000
Archer FusionPlex Sarcoma Panel RNA/TNA Fusions 26 148 1,500,000
Archer FusionPlex Solid Tumor Panel RNA/TNA Fusions/SNVs 53 290 3,000,000
Archer FusionPlex Lung Thyroid Panel RNA/TNA Fusions 8 42 1,500,000
Archer FusionPlex CTL Panel RNA/TNA Fusions/SNVs/Expression 35 195 1,500,000
Archer FusionPlex Oncology Research Panel RNA/TNA Fusions 74 393 3,000,000
Archer FusionPlex ALL Panel RNA/TNA Fusions 81 506 1,500,000
Archer FusionPlex Myeloid Panel RNA/TNA Fusions 84 507 1,500,000
Archer FusionPlex Pan-Heme Panel RNA/TNA Fusions 199 1054 4,500,000

 


 

Why is it better to use RNA instead of DNA for fusion detection?

Why do Archer FusionPlex assays use RNA instead of DNA as input material? It all comes down to biological relevance, cost and turn-around-time. Translocations can occur anywhere in the genome, including introns and other non-coding sequences. They can also occur within the coding regions of genes with limited expression patterns. What this means is that many of the translocations that occur in a cell may not be expressed and thus have little or no biological relevance. For this reason, DNA is not the ideal substrate to search for oncogenic fusions. RNA, on the other hand, is the intermediate product of gene expression and is ideal for detecting fusions, because you are only looking at those that are expressed and potentially oncogenic. Searching for translocations in non-coding regions of the genome is time consuming and expensive. For example, DNA-based hybrid capture techniques tile over intronic regions, which can be repetitive, homopolymer-prone and span 100kb or more. This approach requires more probes, more space on your sequencer and more input material. And even then, coverage can be spotty. On the other hand, FusionPlex assays use RNA transcripts and place gene-specific primers near known fusion breakpoints, so you can identify translocations with a single primer. And because FusionPlex assays combine primers for multiple fusion targets, you can efficiently detect more fusions with less reads and input material. FusionPlex assays use RNA to detect fusions and are better, faster and cheaper than DNA-based hybrid capture techniques. Detect fusions the Archer way, with one of the many FusionPlex assays.

 


 

What values should I expect for cluster density on Illumina® sequencing runs?

The expected cluster density will vary by instrument type and kit version. Please see table below for details.

Instrument Kit Expected Cluster Density
HiSeq® High Output, TruSeq v3 750-850 k/mm2
High Output, TruSeq v4 950-1050 k/mm2
Rapid, v2 850-1000 k/mm2
MiSeq® v2 1000-1200 k/mm2
v3 1200-1400 k/mm2
MiniSeq® Mid & High Output 170-220 k/mm2
NextSeq® Mid & High Output, v2 170-220 k/mm2

Please refer to your kit-specific protocol for complete library denaturation and loading instructions.

 


 

How many Archer® samples can be multiplexed together?

The level of multiplexing depends on the number of targets and the number of reads generated by the instrument per run. This will vary for each catalog panel as well as custom panels. Custom fusion detection assays will need to be optimized to balance the number of reads needed against the level of multiplexing.

Archer® Illumina® or Ion Torrent FusionPlex Panels

Archer Illumina or Ion Torrent Panel Input Material Applications # of Genes # of Targets/Assay Recommended # of Reads
Archer FusionPlex ALK, RET, ROS1 Panel v2 RNA/TNA Fusions/SNVs 3 29 1,000,000
Archer FusionPlex Heme v2 Panel RNA/TNA Fusions 87 607 1,500,000
Archer FusionPlex NTRK Panel RNA/TNA Fusions 3 25 1,000,000
Archer FusionPlex Sarcoma Panel RNA/TNA Fusions 26 148 1,500,000
Archer FusionPlex Solid Tumor Panel RNA/TNA Fusions/SNVs 53 290 3,000,000
Archer FusionPlex Lung Thyroid Panel RNA/TNA Fusions 8 42 1,500,000
Archer FusionPlex CTL Panel RNA/TNA Fusions/SNVs/Expression 35 195 1,500,000
Archer FusionPlex Oncology Research Panel RNA/TNA Fusions 74 393 3,000,000
Archer FusionPlex ALL Panel RNA/TNA Fusions 81 506 1,500,000
Archer FusionPlex Myeloid Panel RNA/TNA Fusions 84 507 1,500,000
Archer FusionPlex Pan-Heme Panel RNA/TNA Fusions 199 1054 4,500,000
Archer FusionPlex Lymphoma Panel RNA/TNA Fusions 125 716 2,000,000

Archer® Illumina® VariantPlex Panels

Archer Illumina Panel # of Targets/Assay Recommended # of Reads
VariantPlex Solid Tumor Panel 660 2,000,000-3,000,00
VariantPlex p53 Panel 23 100,000
VariantPlex CTL Panel 290 1,000,000
VariantPlex Core AML Panel 105 750,000
VariantPlex BRCA1/2 Panel 23 500,000

 


 

What is the average amplicon length for an Archer assay?

FusionPlex Assays

The expected average size for amplicons will range between 150 and 400 base pairs as viewed on a Bioanalyzer trace. However, you should assume an average fragment length of 200 base pairs when using the KAPA Biosystems Library Quantification Kit for qPCR. Our recommended dilutions and MiSeq® and PGM® input amounts are all based on an assumed average fragment length of 200 base pairs. Please refer to the panel specific protocols for guidance.

VariantPlex Assays

The expected average size for amplicons will range between 150 and 400 base pairs as viewed on a Bioanalyzer™ trace. However, assume an average fragment length of 250 base pairs when using the KAPA Biosystems® Library Quantification Kit for qPCR. Our panel-specific recommended dilutions for the MiSeq® and PGM® input amounts are all based on an assumed average fragment length of 250 base pairs. Please refer to the panel specific protocols for guidance.

 


 

What extraction methods are supported for my input material prior to library preparation with the Archer FusionPlex and VariantPlex assays?

For all extraction methods below we recommend the following elution buffer and methods for quantification and QC:

Recommended Elution Buffer Quantification Method Recommended Quality Check Step
nuclease-free water - Qubit® RNA HS assay kit (Life Technologies® Q32852) for FusionPlex
- Qubit® HS dsDNA (Life Technologies® Q32851) for VariantPlex
- Archer® PreSeq RNA QC Assay (AK0043-16) for FusionPlex®
- Archer® PreSeq DNA QC Assay (AK0067-16) for VariantPlex

Total Nucleic Acid Extraction from Formalin-Fixed, Paraffin-Embedded (FFPE) Tissue

Recommended Extraction Method Extraction Kit Protocol Recommendations
Agencourt® FormaPure® Total Nucleic Acid Extraction (A33342) - DO NOT treat with DNase.
- Use heat blocks versus water baths.
- Perform a one-hour digestion at 55°C at Step 5. Do not digest overnight.
- Perform reverse-crosslinking by incubating one hour at 80°C between Steps 5 and 6.
- Elute the sample in 40µL nuclease-free water at Step 23.
Promega ReliaPrep FFPE Total RNA Miniprep System (Z1001) - Extend the reverse crosslinking step at 80°C to one hour.
Maxwell RSC RNA FFPE Kit (AS1440) - Do not DNase treat. Proceed directly from 15 minutes at room temperature to centrifugation at full speed for 2 minutes.

Fresh-Frozen Tissue (FF), Cell Lines and Blood

Recommended Extraction Method
Any total RNA (for FusionPlex) or DNA (for VariantPlex) extraction kit

 


 

What is the benefit of retaining strand specific information in your NGS library?

Tumor specimens are commonly preserved as FFPE samples. Unfortunately formalin fixation can often cause base deamination, resulting in sequencing artifacts. For example, a cytosine on the negative strand is deaminated into a uracil. In traditional opposing primer-based enrichment, the uracil is transcribed into an adenine and the artifact is amplified during PCR. Because amplification occurs before any type of adapters are added to the amplicons, strand specificity is lost, and therefore the sequence analysis will cause a false-positive C to T single nucleotide variant. On the other hand, Anchored Multiplex PCR-based enrichment identifies these deamination events because Molecular Barcode Adapters are ligated to the DNA prior to amplification. Combined with strand- specific primers, AMP maintains the ability to differentiate between positive and negative strand readouts during sequence analysis. So the same C to T transition detected on all negative strands clearly indicates a false-positive SNV, and thus no mutation is called. Let's take a look at actual sequencing data. If this were data from opposing primer-based enrichment, the prevalence of a C to T transition in an FFPE screen would indicate an NRAS G13D variant prevalent in non-small cell lung cancer. But because AMP preserves strand specificity, all of the C to T transitions were detected on the negative strand, demonstrating with extremely high statistical confidence that this was, in fact, an FFPE deamination. Anchored Multiplex PCR is better than traditional opposing primers because strand-specific priming allows you to identify and correct for deamination events that would otherwise lead to false-positive results.

 


 

Can a Qubit® instrument be used to measure the concentrations of the final Archer libraries instead of a KAPA Biosystems® Library Quantification Kit?

Using the Qubit® instrument is not recommended for the final library concentration. We cannot guarantee consistent loading concentrations with the Qubit® because a size selection of the final library is not performed. Therefore, we recommend using the appropriate KAPA Biosystems® Library Quantification Kit for accurate quantification of sequenceable molecules. Other commercially available qPCR-based library quantification kits can also be utilized.

 


 

Is AMP better for novel fusion detection than other library prep methods?

Mutations that drive oncogenesis and disease progression come in all forms, including gene fusions, which can be identified and characterized by sequencing a fusion transcript.

Traditional opposing primer based library preparation methods require target and fusion specific primers that define the region to be sequenced. After amplification, adapters are then ligated to the DNA for further amplification and sequencing. The problem with detecting fusions this way is that you need primers that flank the target region and the fusion partner, so only known fusions can be detected.

Anchored Multiplex PCR enables you to detect the target of interest, plus any known and unknown fusion partners. This is because AMP uses target-specific uni-directional primers, along with reverse primers, that hybridize to the sequencing adapter that is ligated to each fragment prior to amplification. With this approach, your target region, plus any known or novel fusion partners, are selectively amplified for sequencing. This increases the analytical sensitivity of your fusion assay by eliminating false negatives due to novel variants or fusions.

 


 

What are the part numbers for the Archer Molecular Barcode Adapters?

MBC Adapters for Illumina

Catalog # Molecular Barcode Adapter (MBC) Description
SA0040 Archer® MBC Adapters A1-A8 for Illumina®
SA0041 Archer® MBC Adapters A9-A16 for Illumina®
SA0042 Archer® MBC Adapters A17-A24 for Illumina®
SA0043 Archer® MBC Adapters A25-A32 for Illumina®
SA0044 Archer® MBC Adapters A33-A40 for Illumina®
SA0045 Archer® MBC Adapters A41-A48 for Illumina®
AK0016-48 Archer® MBC Adapters Set B for Illumina®
AK0017-48 Archer MBC Adapters Set C for Illumina®

Barcode Adapters for Ion Torrent Platform

Catalog # Molecular Barcode Adapter (MBC) Description
SA0363 Archer® Barcode Adapters 1-8 for Ion Torrent™
SA0364 Archer® Barcode Adapters 9-16 for Ion Torrent™
SA0365 Archer® Barcode Adapters 17-24 for Ion Torrent™
SA0366 Archer® Barcode Adapters 25-32 for Ion Torrent™
SA0367 Archer® Barcode Adapters 33-40 for Ion Torrent™
SA0368 Archer® Barcode Adapters 41-48 for Ion Torrent™


For Research Use Only. Not for use in diagnostic procedures. For Research Use Only. Not for use in diagnostic procedures.

How to contact us

Address

2477 55th Street, Suite 202

Boulder, CO 80301

Phone

Phone: (877) 771 1093

Phone: (303) 357 9001

All content © 2019 ArcherDX, Inc.

For Research Use Only. Not for use in diagnostic procedures. For Research Use Only. Not for use in diagnostic procedures.